The First Automated & Environmentally Controlled Hardness Tester
Long term exposure to temperature and humidity can have drastic effects on rubber hardness readings. The HT 3000 Chamber enables users to designate programmable temperature and humidity changes on test specimens over periods of time. While environmental changes are occurring, 25 samples may be queued for automated data collection in either Shore or IRHD scales – making this suitable for either quality control or R&D applications. Durometer actuation intervals are fully customizable and programmable in conjunction with the chamber conditions. All durometers may also be interchanged using MonTech’s modular durometer system.
Test series on up to 25 sample positions may be identified and grouped while thermal sequences are guided by the software. In Automatic mode, the chamber follows the exact temperature profile, or optional humidity profile, while measurements are taken. This allows long-term aging, or creep test sequences to be executed automatically. For manual sampling, a special sample holder may be placed into the chamber.
Standards | DIN EN ISO 868 | |
Measurement & Technical | ||
Type | Shore A | |
Samples | Cured rubber discs | |
Thickness | 4 mm | |
Spring Force (cN) | 806.50 | |
Contact Pressure / | 12.5 N | |
Indenter | 35° | |
Data Interface | Ethernet (10/100 MBit), USB (int.), CF card (int.), RS232 (optional) | |
Distance (mm) | 2.5 | |
Range | 0 - 100 | |
Shore A Hardness Testing Unit | ||
Power Supply | External | |
Protection Class | IP 50 | |
Reading Accuracy | ± 0.5 SHORE | |
Resolution | 0.1 SHORE | |
Reading Accuracy | ± 0.5 SHORE | |
Pickup Device | ||
Lifting of hardness tester | motorized | |
Load | 10N – optional 12.5 N | |
Indentation Speed | 3,5mm / sec | |
Low Temperature Chamber | ||
Temperature range | -40…180 °C | |
Temperature variation | 0.4…2 +/- °K | |
Temperature fluctuation | 0.1…0.5 +/- °K | |
Average heating-up rate | 4.6 K/min | |
Cooling down time | 93 minutes | |
Max. heat compensation at 25 °C | 500W | |
Interior Volume | 53 | |
Internal Dimensions W x H x D | 24 in x 19 in x 16 in (600 mm x 480 mm x 400 mm) | |
Weight | 230kg | |
Ultra Low Temperature Chamber (optional) | ||
Temperature range | -70…180 °C | |
Temperature variation | 0.2…1.8 +/- °K | |
Temperature fluctuation | 0.1…0.6 +/- °K | |
Average heating-up rate | 5.3 K/min | |
Cooling down time | 9 minutes | |
Max. heat compensation at 25 °C | 1800W | |
Interior volume | 115 | |
Internal Dimensions W x H x D | 24 in x 19 in x 16 in (600 mm x 480 mm x 400 mm) | |
Weight | 385kg | |
Sample Holding Device Embedded in Chamber | ||
Number of Samples | 20 | |
Rotation | Automatic | |
No of test positions per sample | Up to 5 | |
Sample dimensions | Circular buttons | |
Modes | Automatic testing with rotary tray, the rotary tray can easily be uncoupled and a platform (included) can be installed for product testing (e.g. tire segments). | |
Software Features | ||
Connectivity | RS232 and Ethernet | |
Operating System | Win 7, 8 or 10 | |
Computer | optional | |
Operations Modes | Single, series, times, ramp |
The fastest growing rubber testing equipment manufacturer.
13,000 instruments delivered since 1998.
Labs evolve.
So should technology in the instruments.
Speak to real people who understand your account.
Every time.
Service, software and applications specialists are on standby.
And on your time zone.
ISO / IEC 17025 Accredited
ISO 9001:2015 Certified